Keithley 2750 User Manual page 163

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

4-18
Range, Digits, Rate, Bandwidth, and Filter
Figure 4-3
Filter configuration flow chart
SHIFT
TYPE
WINDOW
RDGS
TYPE
Scanning
The moving filter cannot be used when scanning. A scan channel cannot be configured to
use the moving filter. Also, the filter window is not used when scanning.
When a simple scan is configured, the present filter count and state will apply to all chan-
nels in the scan. The window setting is ignored (effectively set to NONE), and if the mov-
ing filter is selected, the filter will not enable when the scan is run.
For the advanced scan, filter state (on or off) and count can be set for each channel. You
cannot set unique filter count, type, and window settings from the advanced scan setup
menu.
For remote programming, the <clist> parameter is used to set filter count and state for
each channel in the scan. You cannot set unique filter type and window settings.
NOTE
0.01%
0.1%
1%
NONE
001 to 100
REPEAT
MOVNG AV
Details on configuring a scan using filtering in a scan are provided in
Model 2750 Multimeter/Switch System User's Manual
Section
7.