External Trigger; Voltmeter Complete - Keithley 2750 User Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System User's Manual

External trigger

The EXT TRIG input requires a falling-edge, TTL-compatible pulse with the specifica-
tions shown in
operations. For the Model 2750 to respond to external triggers, the trigger model must be
configured for it.
Figure 8-4
Trigger link input pulse specifications (EXT TRIG)
TTL High
(2V-5V)
TTL Low
(≤0.8V)

Voltmeter complete

The VMC output provides a TTL-compatible output pulse that can be used to trigger other
instruments. The specifications for this trigger pulse are shown in
you would want the Model 2750 to output a trigger after the settling time of each
measurement.
Figure 8-5
Trigger link output pulse specifications (VMC)
TTL High
(3.4V Typical)
TTL Low
(0.25V Typical)
Figure
8-4. In general, external triggers can be used to control measure
Triggers on
Leading Edge
2μs
Minimum
Meter
Complete
10μs
Minimum
Triggering
8-7
Figure
8-5. Typically,