Channel List Parameter (<Clist>) - Keithley 2750 User Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System User's Manual
Channel list parameter (<clist>)
Channels of one or more switching modules installed in the Model 2750 can be scanned.
Each scan channel can have its own unique setup. For example, a channel could be set to
measure DCV on the 10V range, while another channel can be set to measure ACV on the
1V range.
From the front panel, scan channels are configured from the scan configuration menu as
explained in
ure scan channels.
Channel list parameter:
<clist> = (@SCH)
where: S
Examples:
Throughout this manual, you will encounter commands that can use the <clist> parameter.
The <clist> simply indicates that the associated command can be used to configure a scan
channel. For example:
SENSe:FUNCtion 'VOLTage:AC'
SENSe:FUNCtion 'VOLTage:AC', (@101)
While in the normal measurement display state, the first command simply selects the ACV
function. The second command configures channel 101 to measure ACV when it is
scanned.
See
Section 7
Section
7. For remote programming, the <clist> parameter is used to config-
= Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
for detailed information on scanning.
Basic DMM Operation
' Select ACV function.
' Configure scan channel 101 for ACV.
3-7