Technical data
Repeatability
Installation instructions
Inlet and outlet run
Length of connecting cable
(sensor/transmitter)
60
Measured error at the measuring point
The measured error at the measuring point is made up of the measured error of the device
(0.5 % o.r.) and the measured error resulting from the installation conditions on site. Given a flow
velocity > 0.3 m/s (1 ft/s) and a Reynolds number > 10000, the following are typical error limits:
Nominal diameter
Device error limits
DN 15 (½")
±0.5 % o.r. ± 5 mm/s
DN 25 to 200 (1 to 8")
±0.5 % o.r. ± 7.5 mm/s
> DN 200 (8")
±0.5 % o.r. ± 3 mm/s
o.r. = of reading
Measurement Report
If required, the device can be supplied with a factory measurement report. To certify the
performance of the device, a measurement is performed under reference conditions. Here, the
sensors are mounted on a pipe with a nominal diameter of DN 50 (2") or DN 100 (4") respectively.
The measurement report guarantees the following error limits of the device
[at a flow velocity > 0.3 m/s (1 ft/s) and a Reynolds number > 10000]:
Nominal diameter
DN 50 (2")
DN 100 (4")
o.r. = of reading
±0.3 % for flow velocities > 0.3 m/s (1 ft/s)
10.1.7

Operating conditions: installation

Mounting location
 ä 10
Orientation
 ä 11
 ä 11
The connecting cable is available in the following lengths:
• 5 m (16.4 ft)
• 10 m (32.8 ft)
Installation-specific
+
error limits (typical)
+
±2.5 % o.r.
+
±1.5 % o.r.
+
±1.5 % o.r.
Guaranteed error limits of the device
±0.5 % o.r. ± 5 mm/s
±0.5 % o.r. ± 7.5 mm/s
Proline Prosonic Flow 93T Portable
Error limits at the measuring
point (typical)
 ±3 % o.r. ± 5 mm/s
 ±2 % o.r. ± 7.5 mm/s
 ±2 % o.r. ± 3 mm/s