Proline Promass 83 FOUNDATION Fieldbus
Installation instructions
Inlet and outlet runs
Length of connecting cable
System pressure
Ambient temperature
Storage temperature
Degree of protection
Shock resistance
Vibration resistance
Endress+Hauser
Promass H, I:
DN
8
15
1)
15
25
1)
25
40
1)
40
50
1)
DN 15, 25, 40 "FB" = Full bore versions of Promass I
Promass E:
With nominal diameters DN 8...40, the effect on accuracy of mass flow due to a difference between
calibration pressure and process pressure can be neglected.
With DN 50 the influence is –0.009% o.r. / bar (o.r. = of reading)
Promass A:
A difference between calibration pressure and process pressure has no effect on measuring accuracy.
10.1.7
Operating conditions: Installation
see Page 14 ff.
There are no installation requirements regarding inlet and outlet runs.
Max. 20 meters (remote version)
see Page 15
10.1.8
Operating conditions: Environment
Standard: –20...+60 °C (sensor, transmitter)
Optional: –40...+60 °C (sensor, transmitter)
Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C the readability of the display may be impaired.
–40...+80 °C (preferably +20 °C)
Standard: IP 67 (NEMA 4X) for transmitter and sensor
According to IEC 68-2-31
Acceleration up to 1 g, 10...150 Hz, following IEC 68-2-6
Promass H
% o.r. / bar
–0.017
–0.021
–
–0.013
–
–0.018
–
–0.020
10 Technical data
Promass I
% o.r. / bar
0.006
0.004
0.006
0.006
No influence
No influence
0.006
0.006
99