Technical data
Installation instructions
Connection cable length,
remote version
System pressure
Ambient temperature range
!
Storage temperature
Degree of protection
Shock resistance
Vibration resistance
Electromagnetic compatibility
(EMC)
Medium temperature range
Limiting medium pressure
range
Rupture disk in the sensor
housing
Flow rate
80
11.1.7
Operating conditions: Installation
→ ä 11
Max. 20 m (max. 66 ft)
→ ä 8
11.1.8
Operating conditions: Environment
Sensor and transmitter:
• Standard: –20 to +60 °C (–4 to +140 °F)
• Optional: –40 to +60 °C (–40 to +140 °F)
Note!
• Install the device in a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F), the readability of the display may be impaired.
–40 to +80 °C (–40 to +175 °F), preferably at +20 °C (+68 °F)
Standard: IP 67 (NEMA 4X) for transmitter and sensor
According to IEC 68-2-31
Acceleration up to 2 g, 10 to 150 Hz, following IEC 68-2-6
As per IEC/EN 61326
11.1.9
Operating conditions: Process
–50 to +150 °C (–58 to +302 °F)
Max. 350 bar (5080 psi)
Triggering pressure in the housing 10 to 15 bar (145 to 218 psi)
→ ä 77, "Measuring range"
CNGmass DCI
Endress+Hauser