Operating Conditions: Installation; Suitable Mounting Position - Endress+Hauser Levelflex FMP51 Technical Information

Guided level-radar
Hide thumbs Also See for Levelflex FMP51:
Table of Contents

Advertisement

Suitable mounting position

24
45
40
35
30
25
20
15
10
5
0
500
D
Distance from liquid surface to lower edge of the flange
Delta
Measuring error
If there are fast changes in pressure, there may be an additional error, since the mesured reference distance is
filtered with the time constant of the level measurement. In addition, conditions of imbalance - for example
due to heating - may cause density gradients within the medium and condensation of steam at the probe. As
a result, the level readings at different locations inside the tank may vary slightly. Caused by this application
influences the measuring error may be increased by a factor up to 2 to 3.

Operating conditions: Installation

1
A
C
Lref = 300 mm
1000
1500
2000
D [mm]
2
3
4
Levelflex FMP51, FMP52, FMP54
Lref = 550 mm
2500
3000
3500
4000
B
Endress+Hauser
A0014535
A0012606

Advertisement

Table of Contents
loading

This manual is also suitable for:

Levelflex fmp52Levelflex fmp54

Table of Contents