Proline Prosonic Flow 93T Portable
Endress+Hauser
No.
Error message/type
No. # 6xx Simulation operation active
601
S: POS. ZERO RET.
!: # 601
661
S: SIM. CURR. INP.
to
!: # 661 to 664
664
691
S: SIM. FAILSAFE
!: # 691
692
S: SIM. MEASURAND
!: # 692
698
S: DEV. TEST ACT.
!: # 698
743
S: 0-ADJ.FAIL CHn
to
!: # 743 to 745
745
752
S: W. THICKNESS CH 1
!: # 752
Cause
Positive zero return active.
"
Caution!
This notice message has the highest
display priority!
Current input simulation active.
Simulation of failsafe mode (outputs)
active.
Simulation of a measured variable active
(e.g. mass flow).
The measuring device is being checked on
site via the test and simulation device.
The static zero point calibration of
Channel 1/2 is not possible or was
interrupted.
Channel 1: Wall thickness measurement
active
Troubleshooting
Remedy (spare parts ä 49 ff.)
Switch off positive zero return.
Switch off simulation.
Switch off simulation.
Switch off simulation.
Check that the flow velocity is = 0 m/s.
Switch off wall thickness measurement
53