Preparatory steps prior to installation

Depending on the conditions specific to the measuring point (e.g. Clamp On, number of traverses,
fluid, etc.), a number of preparatory steps have to be taken before actually installing the sensors:
1.
Determination of the values for the necessary installation distances based on the conditions
specific to the measuring point. A number of methods are available for determining the values:
– Local operation of the device
– Applicator (software), online on the Endress+Hauser Internet site
2.
Mechanical preparation of the Clamp On holders for the sensors:
– Mount the sensor holder (DN 15 to 65 / ½ to 2½")
– Premount the strapping bands (DN 50 to 200 / 2 to 8") or (DN 250 to 4000 / 10 to 160")
3.4

Determining the necessary installation distances

The installation distances that have to be maintained depend on:
• The type of sensor: Prosonic Flow P DN 50 to 4000 (2 to 160") or DN 15 to 65 (½ to 2½")
• The type of mounting: Clamp On with strapping band
• Number of traverses or single-path/dual-path version
3.4.1

Installation distances for Prosonic Flow P

DN 50 to 4000 (2 to 160")
1 traverse
SENSOR DISTANCE
WIRE LENGTH
* In the SENSOR DISTANCE function, the distance is indicated in millimeters. The POSITION SENSOR function displays
the values for using the mounting rail (e.g. A3).
3.5

Determining values for installation distances

3.5.1
Determining installation distances via local operation
Perform the following steps to determine the installation distances:
1.
Connect and switch on the transmitter.
2.
Run the "Sensor Installation" Quick Setup menu.
Connecting and switching on the transmitter
1
2
Fig. 7:
Connecting and switching on the transmitter
On/off switch (press switch  3 seconds)
1
2
Charger connection (different adapters are available for the connection)
2 traverses
SENSOR DISTANCE
POSITION SENSOR
mA IN
DATA
12V DC
SERVICE
DN 15 to 65 (½ to 2½")
1 traverse
2 traverses
SENSOR DISTANCE*
POSITION SENSOR*
mA OUT
CH-UP
CH-DN
Installation
A0011547
13