Description of Device Functions
110
11.9.2

Group SYSTEM PARAMETER

Function group CONFIGURATION
BASIC FUNCTION SYSTEM PARAMETER CONFIGURATION
INSTALLATION DIRECTION
SENSOR
(6600)
FLOW DAMPING
(6603)
POSITIVE ZERO RETURN
(6605)
Functional description
Use this function to reverse the sign of the flow quantity, if necessary.
Options
NORMAL
INVERSE
Factory setting
NORMAL
The system damping acts on all functions and outputs of the measuring device.
Use this function to set the filter depth of the digital filter. This reduces the sensitivity
of the measuring signal to interference peaks (e.g. high solids content, gas bubbles in
the fluid, etc.). The system reaction time increases with the filter setting.
User input
0 to 100 s
Factory setting
0 s
Use this function to interrupt evaluation of measured variables. This is necessary when
a piping system is being cleaned, for example. This setting acts on all function and
outputs of the measuring device.
Options
OFF
ON  Signal output is set to the "ZERO FLOW" value.
Factory setting
OFF
Proline Prosonic Flow 93T Portable
Endress+Hauser