11.10 Block Supervision; 11.10.1 Group System - Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual

Portable ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93T Portable:
Table of Contents

Advertisement

Description of Device Functions
114

11.10 Block SUPERVISION

11.10.1 Group SYSTEM

Function group CONFIGURATION
ASSIGN SYSTEM ERROR
(8000)
ERROR CATEGORY
(8001)
ASSIGN PROCESS ERROR
(8002)
ERROR CATEGORY
(8003)
Functional description
SUPERVISION  SYSTEM CONFIGURATION
Use this function to view all system errors and the associated error categories (fault
message or notice message). If you select a single system error you can change
its error category.
User interface
CANCEL
List of system errors with an icon preceding each entry.
!
Note!
• Press the F key twice to call up the function ERROR CATEGORY (8001).
• Use the Q key combination or select "CANCEL" in the system error list to exit the
function.
• A list of possible system errors is provided on  ä 51.
Use this function to define whether a system error triggers a notice message or a fault
message. If you select "FAULT MESSAGES", all outputs respond to an error in
accordance with their defined error response patterns.
Options
NOTICE MESSAGES (display only)
FAULT MESSAGES (outputs and display)
!
Note!
Press the F key twice to call up the function ASSIGN SYSTEM ERROR (8000).
Use this function to view all process errors and the associated error categories (fault
message or notice message). If you select a single process error you can change
its error category.
User interface
CANCEL
List of process errors with an icon preceding each entry.
!
Note!
• Press the F key twice to call up the function ERROR CATEGORY (8003).
• Use the Q key combination or select "CANCEL" in the system error list to exit the
function.
• A list of possible process errors is provided on  ä 54.
Use this function to define whether a process error triggers a notice message or a fault
message. If you select "FAULT MESSAGES", all outputs respond to an error in
accordance with their defined error response patterns.
Options
NOTICE MESSAGES (display only)
FAULT MESSAGES (outputs and display)
!
Note!
Press the F key twice to call up the function
ASSIGN PROCESS ERROR (8002).