Technical data
Power supply failure
Potential equalization
Reference operating
conditions
Maximum measured error
Repeatability
Installation instructions
Inlet and outlet run
Length of connecting cable
112
Lasting min. 1 power cycle:
• EEPROM saves measuring system data if power supply fails
• HistoROM S-DAT: exchangeable data memory which stores the data of the sensor (nominal
diameter, serial number, calibration factor, zero point, etc.)
See Page 53 ff.
10.1.6
Performance characteristics
To DIN EN 29104 and VDI/VDE 2641:
• Fluid temperature: +28 °C ± 2 K
• Ambient temperature: +22 °C ± 2 K
• Warm-up time: 30 minutes
Installation:
• Inlet run >10 x DN
• Outlet run > 5 x DN
• Sensor and transmitter grounded.
• Sensor centered relative to the pipe.
• ± 0.5% o.r. ± 1 mm/s (o.r. = of reading)
• Optional: ± 0.2% o.r. ± 2 mm/s
Supply-voltage fluctuations have no effect within the specified range.
[%]
2.5
2.0
1.5
1.0
0.5
0
0
1
Fig. 60:
Max. measured error in % of reading
Max. ± 0.1% o.r. ± 0.5 mm/s (o.r. = of reading)
10.1.7
Operating conditions: Installation
Any orientation (vertical, horizontal)
Restrictions and additional installation instructions → Page 15 ff.
• Inlet run: typically ≥ 5 x DN
• Outlet run: typically ≥ 2 x DN
For the remote version the permissible cable length L
(→ Page 21, Fig. 17).
A minimum conductivity of 20 µS/cm is required for measuring demineralized water.
0.5 %
0.2 %
2
4
v [m/s]
depends on the conductivity of the medium
max
Proline Promag 50 PROFIBUS DP/PA
6
8
F06-5xxxxxxx-05-xx-xx-xx-000
Endress+Hauser
10