Commissioning; Function Check; Switching On The Measuring Device; Resetting The Measuring Device - Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual

Portable ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93T Portable:
Table of Contents

Advertisement

Commissioning

#
!
40
6
Commissioning
6.1

Function check

Make sure that all final checks have been completed before you commission your measuring point:
• Checklist for "Post-installation check"  ä 30
• Checklist for "Post-connection check"  ä 32
6.2

Switching on the measuring device

Warning!
The measuring device only complies with the general safety requirements in accordance with
EN 61010-1 and the EMC requirements of IEC/EN 61326 during storage battery operation.
Disconnect the charger from the measuring device for measuring operation.
The measuring device is switched on by pressing the ON/OFF switch  3 seconds ( ä 31,
 å 36, No. 1).
The measuring system performs a number of internal test functions after power-up. During this
process, the following sequence of messages appears on the local display:
PROSONIC FLOW 93
START-UP
RUNNING
Æ
PROSONIC FLOW 93
DEVICE SOFTWARE
V XX.XX.XX
Æ
CURRENT OUTPUT
...
Æ
SYSTEM OK
OPERATION
Æ
Normal measuring mode commences as soon as startup completes.
Various measured value and/or status variables appear on the display (HOME position).
Note!
If startup fails, an appropriate error message is displayed, depending on the cause.
6.2.1

Resetting the measuring device

The measuring device is reset by pressing the ON/OFF switch  30 seconds ( ä 31,  å 36,
No. 1). Only the internal clock of the measuring device is reset during the reset. All the other
settings remain unchanged.
Startup message
Display of current software
Display of installed input/output modules
Beginning of standard measuring mode
Proline Prosonic Flow 93T Portable
Endress+Hauser

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents