Block Totalizer; Group Totalizer (1 To 3) - Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual

Portable ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93T Portable:
Table of Contents

Advertisement

Proline Prosonic Flow 93T Portable
Endress+Hauser
11.6

Block TOTALIZER

11.6.1

Group TOTALIZER (1 to 3)

Function group CONFIGURATION
ASSIGN
(3000)
UNIT TOTALIZER
(3001)
TOTALIZER MODE
(3002)
RESET TOTALIZER
(3003)
Functional description
TOTALIZER TOTALIZER  CONFIGURATION
Use this function to assign a measured variable to the totalizer in question.
Options
OFF
VOLUME FLOW (CH1)
Factory setting
VOLUME FLOW (CH1)
!
Note!
• The totalizer is reset to 0 as soon as the selection is changed.
• If you select OFF in the function group CONFIGURATION of the totalizer in
question, only the ASSIGN (3000) function remains visible.
Use this function to define the unit for the totalizer's measured variable, as selected
beforehand.
Options
Metric:
cm3; dm3; m3; ml; l; hl; Ml
US:
cc; af; ft3; oz f; gal; Mgal; bbl (normal fluids); bbl (beer); bbl (petrochemicals);
bbl (filling tanks)
Imperial:
gal; Mgal; bbl (beer); bbl (petrochemicals)
Arbitrary unit (from ARBITRARY UNIT function group  ä 71):
_ _ _ _
Factory setting
3
m
Use this function to define how the flow components are to be totaled by the totalizer
in question.
Options
BALANCE
Positive and negative flow components. The positive and negative flow components
are balanced. In other words, net flow in the flow direction is registered.
FORWARD
Positive flow components only
REVERSE
Negative flow components only
Factory setting
Totalizer 1 = BALANCE
Totalizer 2 = FORWARD
Totalizer 3 = REVERSE
Use this function to reset the sum and the overflow of the totalizer in question to zero.
Options
NO
YES
Factory setting
NO
Description of Device Functions
85