Block Inputs; Group Current Input - Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual

Portable ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93T Portable:
Table of Contents

Advertisement

Description of Device Functions
100
11.8

Block INPUTS

11.8.1

Group CURRENT INPUT

Function group CONFIGURATION
ASSIGN
(5200)
CURRENT SPAN (5201)
VALUE 0-4 mA (5202)
VALUE 20 mA
(5203)
ERROR VALUE
(5204)
Functional description
INPUTS CURRENT INPUT CONFIGURATION
Use this function to assign a process variable to the current input.
Options
VOLUME FLOW DEVICE 2
OFF
Factory setting
OFF
Use this function to define the current span. The selection specifies the operational
range and the lower and upper signal on alarm.
Options
0–20 mA
4–20 mA
4–20 mA NAMUR
4–20 mA US
0–20 mA (25 mA)
4–20 mA (25 mA)
Factory setting
4–20 mA NAMUR
Current range / operational range (measuring information)
0–20 mA / 0 to 20.5 mA
4–20 mA / 4 to 20.5 mA
4–20 mA NAMUR / 3.8 to 20.5 mA
4–20 mA US / 3.9 to 20.8 mA
0–20 mA (25 mA) / 0 to 24 mA
4–20 mA (25 mA) / 4 to 24 mA
Use this function to assign the 0/4 mA current a value.
Options
5-digit floating-point number
Factory setting
3
0 m
/h
Use this function to assign the 20 mA current a value.
Options
5-digit floating-point number
Factory setting
10 l/s
Use this function to enter a defined error value for the process variable in question.
!
Note!
If the current value is outside the selected range (see CURRENT SPAN function,
(5201)), the process variable is set to the "error value" defined here and a
corresponding notice message "RANGE CUR. IN
(# 363)" is generated.
User input
5-digit floating-point number
Factory setting
0 l/s
Proline Prosonic Flow 93T Portable
Endress+Hauser

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents