Operating Conditions: Installation; Operating Conditions: Environment - Endress+Hauser Proline Promass 84 MODBUS RS485 Operating Instructions Manual

Coriolis mass flow measuring system for custody transfer
Table of Contents

Advertisement

Technical data
Installation instructions
Inlet and outlet runs
Connection cable length,
remote version
System pressure
Ambient temperature range
!
Storage temperature
Ambient class
Degree of protection
Shock resistance
Vibration resistance
CIP cleaning
SIP cleaning
Electromagnetic compatibility
(EMC)
106
11.1.7

Operating conditions: Installation

→ ä 12
There are no installation requirements regarding inlet and outlet runs.
max. 20 m (65 ft)
→ ä 13
11.1.8

Operating conditions: Environment

Sensor and transmitter:
• Standard: –20 to +60 °C (–4 to +140 °F)
• Optional: –40 to +60 °C (–40 to +140 °F)
Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.
–40 to +80 °C (–40 to +175 °F), preferably at +20 °C (+68 °F)
B, C, I
Standard: IP 67 (NEMA 4X) for transmitter and sensor
According to IEC 68-2-31
Acceleration up to 1 g, 10 to 150 Hz, following IEC 68-2-6
yes
yes
To IEC/EN 61326 and NAMUR recommendation NE 21
Proline Promass 84 MODBUS RS485
Endress+Hauser