Block Outputs; Group Current Output - Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual

Portable ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93T Portable:
Table of Contents

Advertisement

Proline Prosonic Flow 93T Portable
Endress+Hauser
11.7

Block OUTPUTS

11.7.1

Group CURRENT OUTPUT

Function group CONFIGURATION
OUTPUTS E  CURRENT OUTPUT 1 EAA  CONFIGURATION 400
ASSIGN CURRENT
OUTPUT
(4000)
Functional description
Use this function to define a second reading to be displayed in the main line alternately
(every 10 seconds) with the reading defined in the function ASSIGN (2600).
Options
OFF
VOLUME FLOW CH 1
SOUND VELOCITY CH 1
SIGNAL STRENGTH CH 1
FLOW VELOCITY CH 1
Factory setting
VOLUME FLOW CH 1
!
Note!
If you select OFF, the only function shown in the function group CONFIGURATION
(400) is this function, in other words, ASSIGN CURRENT OUTPUT (4000).
Description of Device Functions
87