Troubleshooting
!
56
9.5
Response of outputs to errors
Note!
The response of the totalizer (failsafe mode) can be configured by means of various functions in the
function matrix.
You can use positive zero return to set the signals to their fallback value, for example when
operation has to be interrupted while a pipe is being cleaned. This function has priority over all other
device functions; simulations are suppressed, for example.
Error response mode of outputs and totalizers
Process/system error present
"
Caution!
System or process errors defined as "notice messages" have no effect whatsoever on the inputs and outputs.
See the information on ä 38 ff.
Totalizer
STOP
The totalizers stop if a fault is present.
ACTUAL VALUE
The fault is ignored. The totalizers continue to count in accordance with the
current flow measured value.
HOLD VALUE
The totalizers continue to count in accordance with the last valid flow measured
value (before the fault occurred).
9.6
Spare parts
Only accessories are available for the measuring device ä 49.
9.7
Return
ä 6
9.8
Disposal
Observe the regulations applicable in your country!
9.9
Software history
Date
Software version
06.2011
2.03.XX
2.02.XX
04.2010
06.2009
2.02.XX
Changes to software
– Added Site Manager
– Added Current Output
– Added Liner Information to Quick Set-up
– Added some ANSI standards to pipe standards
No changes to software
Original software for Prosonic Flow 93T Portable
Proline Prosonic Flow 93T Portable
Positive zero
return activated
The totalizer
stops.
Operating
Instructions
71136630/13.11
71112144/04.10
71093720/06.09