Commissioning
46
6.7

Site Manager

The Prosonic Flow 93T is equipped with the capability to store and recall user information for a
given set-up. This set of programmed (such as Pipe Data, Sensor Parameters, Liquid Data, etc.)
defines a specific a "site". Using the Site Manager feature the user can store site data in files for
current and future access.
One site is in always stored in the instrument's internal memory. The storage of additional sites is
physically located on the USB stick. Thus, the Site Manager can only be accessed and used when
the USB stick is installed in the Prosonic Flow 93T transmitter. A maximum of 20 sites can be cre-
ated and stored.
The Site Manager is accessed via the Output function block. OUTPUTS  F DATA LOGGER F 
SITE MANAGER.
The Site Manager allows the user to perform the following specific functions:
• SAVE a set of programmed information as a site to the USB stick  ä 98
• Assign a user defined NAME to the site  ä 98
• LOAD a site stored on the USB stick to the 93T transmitter  ä 98
• DELETE sites stored on the USB stick  ä 99
The "site" data currently being used by the instrument for measurement (the data in the instrument's
internal memory) is referred to as the ACTUAL SITE. When the Site Manager is first accessed the
instrument compares the site data currently being used in its internal memory with the all the site
data files located on the external USB stick. During this process the meter will display the message
CHECKING. If there is a match between these two data sets the unit will display the name of the
site in the ACTUAL SITE position. If there is no match the instrument will display "_ _ _ _ _ _".
The user then has ability to save this data as a NEW SITE and assign a name or choose a previously
stored site and LOAD that site to the instrument.
A site is defined only by the most important information specific to a given metering application, the
table below indicates which specific information is stored when a site is created or loaded into the
instrument memory.
Group SYSTEM UNITS
Group SENSOR PARAMETER
Group PIPE DATA
UNIT VOLUME FLOW (0042)
UNIT TEMPERATURE (0422)
UNIT VISCOSITY (0423)
UNIT LENGTH (0424)
UNIT VELOCITY (0425)
MEASUREMENT (6880)
SENSOR TYPE (6681)
SENSOR CONFIGURATION (6882)
PIPE STANDARD (6520)
NOMINAL DIAMETER (6521)
PIPE DIAMETER (6526)
PIPE MATERIAL (6522)
SOUND VELOCITY PIPE (6524)
WALL THICKNESS (6527)
LINER MATERIAL (6528)
SOUND VELOCITY LINER (6529)
LINER THICKNESS (6528)
Proline Prosonic Flow 93T Portable