Mounting with strapping bands (flexible)

For sensor Prosonic Flow P (DN 50 to 4000 / 2 to 160")
Caution!
• Each time you use the strapping bands, check that the ratchet locks and springs function safely
beforehand.
• Inspect the strapping bands for damage.
Procedure
Closing the strapping band lock
1.
Fit the mounting bolt onto the strapping band.
2.
Guide the strapping band around the pipe making sure it is not twisted in the process and, with
the ratchet lock (a) open, push the end through the slot. Pretension manually by pulling on the
free end of the strapping band.
!
Note!
If you do not pretension the bands it is more difficult to release the strapping bands.
3.
Tension continuously by moving the lever back and forth (b) until the strapping band is
optimally tensioned.
4.
Then push down the lever (c).
"
Caution!
The tensioning clamp (d) must engage on both sides!
Opening the strapping band lock
1.
Pull back the lever lock (e) while simultaneously opening the lever 180° (f) until the lever lock
(g) is engaged.
2.
Remove the strapping band.
a
Fig. 18:
Strapping band lock
1
Closing the strapping band lock
2
Opening the strapping band lock
f e
c
b
d
1
Proline Prosonic Flow 93T Portable
g
g
2
Endress+Hauser
A0011556.