Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual page 115

Portable ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93T Portable:
Table of Contents

Advertisement

Proline Prosonic Flow 93T Portable
Endress+Hauser
ACKNOWLEDGE FAULT
(8004)
ALARM DELAY
(8005)
Function group OPERATION
ACTUAL SYSTEM
CONDITION
(8040)
PREVIOUS SYSTEM
CONDITIONS
(8041)
SIMULATION FAILSAFE
MODE (8042)
Functional description
SUPERVISION  SYSTEM CONFIGURATION
Use this function to define the measuring device's response to fault messages.
Options
OFF
The measuring device resumes normal operation when the fault is rectified. The fault
message disappears automatically.
ON
The measuring device resumes normal operation when the fault is rectified. The fault
message has to be acknowledged by pressing the F key on the local display.
Factory setting
OFF
Use this function to specify a time period for suppressing the appearance of fault or
notice messages.
Depending on the setting and the type of error, this suppression acts on:
• Display
• Relay output
• Current output
• Frequency output
User input
0 to 100 s (in steps of one second)
Factory setting
0 s
"
Caution!
If this function is activated, error and notice messages are delayed by the time
corresponding to the setting before being forwarded to the higher-order controller
(process controller, etc.).
It is therefore imperative to check in advance in order to make sure whether a delay of
this nature could affect the safety requirements of the process. If error and notice
messages cannot be suppressed, a value of 0 seconds must be entered here.
Functional description
SUPERVISION  SYSTEM OPERATION
Use this function to check the present system condition.
User interface
"SYSTEM OK" or the fault / notice message with the highest priority.
Use this function to view the fifteen most recent error and notice messages since
measuring last started.
User interface
The last 15 fault/notice messages appear on the display.
Use this function to set all inputs, outputs and totalizers to their defined failsafe modes,
in order to check whether they respond correctly. During this time, the words
"SIMULATION FAILSAFE MODE" appear on the display.
Options
ON
OFF
FAILURE (CH1)
Factory setting
OFF
Description of Device Functions
115