10 Technical data
Installation instructions
Inlet and outlet runs
Length of connecting cable
System pressure
Ambient temperature
Storage temperature
Degree of protection
Shock resistance
Vibration resistance
Suitability for CIP cleaning
Suitability for SIP cleaning
Electromagnetic compatibility
(EMC)
104
10.1.7

Operating conditions: Installation

see Page 14 ff.
There are no installation requirements regarding inlet and outlet runs.
Max. 20 meters (remote version)
see Page 15.
10.1.8

Operating conditions: Environment

Standard: –20 °C to +60 °C (sensor, transmitter)
Optional: –40 °C to +60 °C (sensor, transmitter)
Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C the readability of the display may be impaired.
–40 °C to +80 °C (preferably +20 °C)
Standard: IP 67 (NEMA 4X) for transmitter and sensor
According to IEC 68-2-31
Acceleration up to 1 g, 10 to 150 Hz, following IEC 68-2-6
yes
yes
To EN 61326/A1 and NAMUR recommendation NE 21/43
Proline Promass 80 PROFIBUS PA
Endress+Hauser