Installing Prosonic Flow P (Dn 15 To 65 / ½ To 2½"); Mounting The Sensor; Installing Prosonic Flow P - Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual

Portable ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93T Portable:
Table of Contents

Advertisement

Proline Prosonic Flow 93T Portable
!
Endress+Hauser
3.7
Installing Prosonic Flow P (DN 15 to 65 / ½ to 2½")
3.7.1

Mounting the sensor

Prerequisites
• The sensor holder is already mounted  ä 16.
• The distance of the sensor holder is set (sensor distance)  ä 13.
Material
The following material is needed for mounting:
• Sensor
• Connecting cable
Note!
Prior to mounting, connect the connecting cables to the sensors.
Procedure
1.
Coat the contact surfaces (1) of the sensors with an even layer of coupling fluid
approx. 1 mm (0.04") thick.
Fig. 19:
Coating with coupling fluid
2.
Mount as illustrated in the graphic (steps 1 to 5):
1
4
Fig. 20:
Mounting the sensors,
Sensor holder: Model 1
This completes the mounting process. The sensors can now be connected to the transmitter via the
connecting cables  ä 31.
1
2
1
3
5
4
A0011552
Fig. 21:
Mounting the sensors,
Sensor holder: Model 2
Installation