3.10 Installing Prosonic Flow P (DN 50 to 4000 / 2 to 160")

3.10.1 Installation for measurement via one traverse
!
Note!
The procedure is identical for Prosonic Flow W.
Prerequisites
• The installation distances (sensor distance and wire length) are known  ä 8.
• The strapping bands are already mounted  ä 16/ ä 17.
Material
The following material is needed for mounting:
• Two strapping bands incl. mounting bolts and centering plates where necessary (already
mounted )
• Two measuring wires, each with a cable lug and a fixer to position the strapping bands
• Two sensor holders
• Coupling fluid for an acoustic connection between the sensor and pipe
• Two sensors incl. connecting cables.
Procedure
1.
Prepare the two measuring wires:
– Arrange the cable lugs and fixer such that the distance they are apart corresponds to the
wire length (SL).
– Screw the fixer onto the measuring wire.
Fig. 12: Fixer (a) and cable lugs (b) at a distance that corresponds to the wire length (SL)
2.
With the first measuring wire:
– Fit the fixer over the mounting bolt of the strapping band that is already securely
mounted.
– Run the measuring wire clockwise around the pipe.
– Fit the cable lug over the mounting bolt of the strapping band that can still be moved.
3.
With the second measuring wire:
– Fit the cable lug over the mounting bolt of the strapping band that is already securely
mounted.
– Run the measuring wire counterclockwise around the pipe.
– Fit the fixer over the mounting bolt of the strapping band that can still be moved.
Endress+Hauser
SL
1
2
3
4
5
6
7
8
9
10
11
Installation
A0001112
21