Raman flow assembly
5

Installation

5.1

Installing the micro flow bench

The Endress+Hauser micro flow bench slips onto the Rxn-10 probe and is secured by a torque limiting thumb screw
clamp. The thumb screw on the Rxn-10 probe should never be fully removed.
When installing or removing optics, ensure the laser and emission shutter are in the closed position.
To install the Rxn-10 probe onto the micro flow bench:
1. Loosen the metal screw on the Rxn-10 probe by turning the screw counter-clockwise approximately one
turn (do not remove).
2. Insert the end optic clamp of the probe onto the Rxn-10 adapter of the micro flow bench.
Figure 1. Inserting the Rxn-10 probe onto the Rxn-10 adapter of the micro flow bench
3. Slide the probe over the Rxn-10 adapter of the micro flow bench until it stops.
4. Tighten the thumb screw by gently turning it clockwise until there is an audible "click" sound. This indicates
the thumb screw has reached the desired torque. Failure to tighten the screw will result in the optic coming
loose, potentially damaging the optic.
5. After installing the micro flow bench, use the micro flow bench calibration kit to perform an intensity
calibration for the probe with the new optic.
To remove the Rxn-10 probe from the micro flow bench:
1. Loosen the torque limiting thumb screw by turning it counter-clockwise approximately one turn so that the
Rxn-10 adapter is released from the clamp. Do not remove the screw.
2. Slide the probe off of the adapter.
Endress+Hauser
Figure 2. Final position of the Rxn-10 probe with the micro flow bench
Operating Instructions
A0052579
A0052580