Installing The Measuring Sensors Prosonic Flow W, P - Endress+Hauser prosonic flow 92 Operating Instructions Manual

Portable ultrasonic flow measuring system
Hide thumbs Also See for prosonic flow 92:
Table of Contents

Advertisement

3 Installation
20
3.3.2

Installing the measuring sensors Prosonic Flow W, P

1 traverse version
1.
Fix a tensioning band for small or large nominal diameters as described on Page 17
and Page 18.
Install the second tensioning band (threaded bolt on the opposite side). The sec-
ond tensioning band must still be moveable.
Intermediate step
The wire length is determined via the "Site Setup" menu (see Page 42, 68).
2.
Enter the wire length on both halves of the wire.
Fig. 12:
Marking off the determined wire length on the wire measurement equipment (SL = wire length)
3.
Push the cable lug and the fixer over the first threaded bolt. Lead each wire along
one side of the pipe. Push the cable lug and the fixer over the second threaded bolt.
Pull in the threaded bolt with the tensioning band until both cords are the same
length. Fix the tensioning band. Loosen the Phillips screws of the fixing parts.
Remove the cords.
Fig. 13:
Use of the wire measuring equipment for positioning the threaded bolts
SL
1
2
3
4
5
6