2.7.1 Mounting the sensor
Prerequisites
• The sensor holder is already mounted → ä 11.
• The distance of the sensor holder is set (sensor distance) → ä 11.
Material
The following material is needed for mounting:
• Sensor
• Connecting cable
!
Note!
Prior to mounting, connect the connecting cables to the sensors.
Procedure
1.
Coat the contact surfaces (1) of the sensors with an even layer of coupling fluid
approx. 1 mm (0.04") thick.
Fig. 11: Coating with coupling fluid
Endress+Hauser
1
Installation
A0013624
17