Installing Prosonic Flow W And Pdn 15 To 65 (½ To 2½"); Mounting The Sensor - Endress+Hauser Proline Prosonic Flow 93 Operating Instructions Manual

Foundation fieldbus ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93:
Table of Contents

Advertisement

Proline Prosonic Flow 93 FOUNDATION Fieldbus
Endress+Hauser
3.8
Installing Prosonic Flow W and P DN 15 to 65
(½ to 2½")
3.8.1

Mounting the sensor

Prerequisites
• The installation distance (sensor distance) is known  ä 16.
• The sensor holder is already mounted  ä 30.
Material
The following material is needed for mounting:
• Sensor incl. adapter cable
• Connecting cable for connecting to the transmitter
• Coupling fluid for an acoustic connection between the sensor and pipe
Procedure
1.
Set the distance between the sensors as per the value determined for the sensor distance. Press
the sensor down slightly to move it.
Fig. 25:
Setting the distance between the sensors as per the value for the sensor distance
a
Sensor distance
b
Contact surfaces of the sensor
2.
Coat the contact surfaces of the sensors with an even layer of coupling fluid
(approx. 0.5 to 1 mm / 0.02 to 0.04") thick.
3.
Fit the sensor housing on the sensor holder.
Fig. 26:
Fitting the sensor housing
Installation
b
A0011529
A0011527
35

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Proline prosonic flow pProline prosonic flow w

Table of Contents