3 Installation
20
3.3.3

Installing the measuring sensors Prosonic Flow P

2 or 4 traverses version
1.
Fix a tensioning band for small or large nominal diameters as described on
Page 17.
2.
Fix the sensor holder to the pipe using the threaded stud. Tighten the locking nut
using a spanner (AF 13).
3.
Fasten the mounting rail bracket to the sensor holder using a Philips screwdriver.
Fig. 12:
Installation steps 1 to 3, measuring sensors Prosonic Flow P, 2 or 4 traverses version
Intermediate step:
Install the second sensor holder as per steps 1 to 3. Make sure that the tensioning band
is not fastened. This means the sensor holder can still be moved along the pipe.
4.
Determine the spacing between the distance holes (sensor distance).
!
Note!
To determine the spacing between the distance holes (sensor distance) use:
– the "Sensor Installation" Quick Setup menu if the measuring device has a local
operation. Run the Quick Setup as described on Page 82. The sensor distance is
displayed in the POSITION SENSOR function (6884) (i.e. a letter between A...K
for sensor 1 and a number between 10...76 for sensor 2). Insert the mounting rail
as appropriate (for DN 50...200 or for DN 200...600) and tighten the relevant
screws. Now fasten the second tensioning band. The transmitter has to be
installed and connected to the power supply to carry out the "Sensor Installation"
Quick Setup.
– the procedure described on Page 86 ff. if the measuring device has no local
operation.