Technical data

Measuring principle
Measuring system
Measured variable
Measuring range
130
16
Technical data
16.1
Application
The measuring device is suitable for flow measurement of gases only.
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
To ensure that the device remains in proper operating condition for its service life, use the
measuring device only for media against which the process-wetted materials are
adequately resistant.
16.2
Function and system design
Proline Prosonic Flow uses a measurement method based on the transit time difference.
The device consists of a transmitter and a sensor.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
For information on the structure of the device →  12
16.3
Input
Direct measured variables
Volume flow
Calculated measured variables
• Corrected volume flow
• Mass flow
Optional measured variables (can be ordered)
Order code for "Sensor version", option 2 "Volume flow + Biogas analysis"
• Corrected methane volume flow
• Energy flow
• Methane fraction
• Gross calorific value
• Wobbe index
• Temperature
Standard (order code for "Calibration flow", option 1 "Operable flow range 30 : 1")
Nominal diameter
[mm]
[in]
[m/s]
50
2
1 to 30
80
3
1 to 30
100
4
1 to 30
Velocity
[ft/s]
[m
3.28 to 98.4
9 to 269
3.28 to 98.4
20 to 611
3.28 to 98.4
34 to 1 032
Proline Prosonic Flow B 200 HART
Effective volume flow
3
3
/h]
[ft
/h]
316 to 9 495
720 to 21 592
1 215 to 36 443
Endress+Hauser