Endress+Hauser Proline Prosonic Flow B 200 Operating Instructions Manual page 138

Ultrasonic transit time flowmeter
Hide thumbs Also See for Proline Prosonic Flow B 200:
Table of Contents

Advertisement

Technical data
Ambient temperature
range
Storage temperature
Degree of protection
Shock resistance
Vibration resistance
138
Current output
Additional error, in relation to the span of 16 mA:
Temperature coefficient at
0.02 %/10 K
zero point (4 mA)
Temperature coefficient
0.05 %/10 K
with span (20 mA)
Pulse/frequency output
Temperature coefficient
Max. ±100 ppm o.r.
16.7
Installation
"Mounting requirements" →  18
16.8
Environment
Transmitter
Local display
Sensor
If operating outdoors:
Avoid direct sunlight, particularly in warm climatic regions.
All components apart from the display modules:
–40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F)
Display modules
–40 to +80 °C (–40 to +176 °F)
Transmitter
• As standard: IP66/67, type 4X enclosure
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
Sensor
IP66/67, type 4X enclosure
In accordance with EN 60721-3-4
Class 4M4, in accordance with EN 60721-3-4
–40 to +60 °C (–40 to +140 °F)
–20 to +60 °C (–4 to +140 °F), the readability of the display may be
impaired at temperatures outside the temperature range.
• Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)
• Flange material stainless steel: –40 to +60 °C (–40 to +140 °F)
• Version without flange: –40 to +60 °C (–40 to +140 °F)