A maximum of 20 event messages can be displayed in chronological order. If the advanced
HistoROM function is enabled in the device (order option), up to 100 entries can be
displayed.
The event history includes entries for:
• Diagnostic events →  112
• Information events →  117
In addition to the operation time of its occurrence, each event is also assigned a symbol
that indicates whether the event has occurred or is ended:
• Diagnostic event
: Event has occurred
: Event has ended
• Information event
: Event has occurred
To call up the measures to rectify a diagnostic event:
• Via local display →  109
• Via "FieldCare" operating tool →  111
For filtering the displayed event messages →  117
12.8.2
Filtering the event logbook
Using the Filter options parameter, you can define which category of event messages is
displayed in the Events list submenu.
Navigation path
"Diagnostics" menu → Event logbook → Filter options
Filter categories
• All
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
• Information (I)
12.8.3
Overview of information events
Unlike a diagnostic event, an information event is displayed in the event logbook only and
not in the diagnostic list.
Info number
I1000
--------(Device ok)
I1079
Sensor changed
I1089
Power on
I1090
Configuration reset
I1091
Configuration changed
I1092
Trend data deleted
/ ../Eventlist
I1091 Config. change
I1157 Mem.err. ev.list
0d01h19m10s
F311 Electr. failure
Info name
Diagnostics and troubleshooting
F
A0014008-EN
117