Configurations - Endress+Hauser Proline Prosonic Flow B 200 Operating Instructions Manual

Ultrasonic transit time flowmeter
Hide thumbs Also See for Proline Prosonic Flow B 200:
Table of Contents

Advertisement

Proline Prosonic Flow B 200 HART
Endress+Hauser
Diagnostic
Short text
number
841
Sensor range
842
Process limit
881
Sensor signal path 1 to
2
882
Input signal
930
Process fluid
931
Process fluid
1)
Diagnostic behavior can be changed.
12.6
Pending diagnostic events
The Diagnostics menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
• Via local display →  109
• Via "FieldCare" operating tool →  111
Other pending diagnostic events can be displayed in the Diagnostic list submenu
→  116
Navigation
"Diagnostics" menu
Structure of the submenu
Diagnostics
Diagnostics and troubleshooting
Remedy instructions
Check flow velocity
Low flow cut off active!
1. Check low flow cut off
configuration
1. Check process conditions
2. Clean transducer
3. Change transducer
1. Check input configuration
2. Check external device or process
conditions
Sound velocity too high! Check process
conditions
Sound velocity too low! Check process
conditions
Actual diagnostics
Previous diagnostics
Operating time from restart
Operating time
Status
Diagnostic
signal
behavior
[from the
[from the
factory]
factory]
1)
S
Warning
S
Warning
1)
M
Warning
F
Alarm
S
Warning
S
Warning
115