Endress+Hauser Proline Prosonic Flow B 200 Operating Instructions Manual page 137

Ultrasonic transit time flowmeter
Hide thumbs Also See for Proline Prosonic Flow B 200:
Table of Contents

Advertisement

Proline Prosonic Flow B 200 HART
Repeatability
Response time
Influence of ambient
temperature
Endress+Hauser
Example for max. measured error (volume flow)
[%]
10.0
9.0
8.0
7.0
6.0
5.0
4.0
3.0
2.0
1.0
0
0
1
2
0
5
 35
Example for max. measured error (volume flow) in % o.r.
1
Standard (order code for "Calibration flow", option 1 "Operable flow range 30 : 1")
2
Optional (order code for "Calibration flow", option 2 "Operable flow range 100 : 1")
Accuracy of outputs
o.r. = of reading
The outputs have the following base accuracy specifications.
Current output
Accuracy
±10 µA
Pulse/frequency output
Accuracy
Max. ±100 ppm o.r.
o.r. = of reading; o.f.s. = of full scale value; abs. = absolute; T = medium temperature
Volume flow
±0.5 % o.r.
Methane
±0.5 % o.f.s. = ±0.5 % abs.
Temperature
±0.3 °C ± 0.0025 × T °C (±0.45 °F ± 0.0025 × (T – 32) °F)
• The response time depends on the configuration (damping).
• Response time in the event of erratic changes in the flow: after 1 000 ms 95% of the full
scale value.
o.r. = of reading
2
1
3
4
5
6
20
10
15
Technical data
7
8
30
25
100
[m/s]
v
[ft/s]
A0015541
137