Proline Prosonic Flow B 200 HART
Endress+Hauser
Diagnostic
Short text
number
123
Predicted signal
strength
124
Relative signal
strength
125
Relative sound velocity 1. Clean the transducer
160
Signal path switched
off
Diagnostic of electronic
242
Software incompatible
252
Modules incompatible
261
Electronic modules
262
Module connection
270
Main electronic failure Change main electronic module
271
Main electronic failure 1. Restart device
272
Main electronic failure 1. Restart device
273
Main electronic failure 1. Emergency operation via display
275
I/O module failure
276
I/O module failure
282
Data storage
283
Memory content
302
Device verification
active
311
Electronic failure
311
Electronic failure
384
Transmitter circuit
385
Amplifier circuit
386
Time of flight
Diagnostic of configuration
410
Data transfer
Diagnostics and troubleshooting
Remedy instructions
1. Check parameterization in the
' M edium' menu
2. Check process conditions
3. Clean the transducer
4. Replace the transducer
1. Clean the transducer
2. Replace the transducer
2. Replace the transducer
Contact service
1. Check software
2. Flash or change main electronics
module
1. Check electronic modules
2. Change I/O or main electronic
module
1. Restart device
2. Check electronic modules
3. Change I/O Modul or main
electronics
1. Check module connections
2. Change electronic modules
2. Change main electronic module
2. Contact service
2. Change main electronics
Change I/O module
1. Restart device
2. Change I/O module
1. Restart device
2. Contact service
1. Transfer data or reset device
2. Contact service
Device verification active, please wait.
1. Transfer data or reset device
2. Contact service
Maintenance required!
1. Do not perform reset
2. Contact service
Change main electronic module
Change main electronic module
Change main electronic module
1. Check connection
2. Retry data transfer
Status
Diagnostic
signal
behavior
[from the
[from the
factory]
factory]
1)
M
Warning
1)
M
Warning
1)
M
Warning
M
Warning
F
Alarm
1)
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
C
Warning
F
Alarm
M
Warning
F
Alarm
F
Alarm
F
Alarm
F
Alarm
113