Proline Prosonic Flow B 200 HART
Parameter
Operating time
Last backup
Configuration management
Comparison result
Endress+Hauser
Parameter overview with brief description
Prerequsite
A local display is provided.
A local display is provided.
A local display is provided.
10.6.1
Function scope of the "Configuration management" parameter
Options
Description
Execute backup
The current device configuration is backed up from the integrated HistoROM to the
device' s display module. The backup copy includes the transmitter data of the
device.
Restore
The last backup copy of the device configuration is restored from the display
module to the device' s integrated HistoROM. The backup copy includes the
transmitter data of the device.
Duplicate
The transmitter configuration from another device is duplicated to the device using
the display module.
Compare
The device configuration saved in the display module is compared with the current
device configuration of the integrated HistoROM.
Clear backup data
The backup copy of the device configuration is deleted from the display module of
the device.
Integrated HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
10.7
Simulation
The Simulation submenu enables you to simulate, without a real flow situation, various
process variables in the process and the device alarm mode and to verify downstream
signal chains (switching valves or closed-control loops).
Description
Indicates how long the device
has been in operation.
Indicates when the last data
backup was saved to the
display module.
Select action for managing the
device data in the display
module.
Comparison between present
device data and display
backup.
Commissioning
User interface / 
Factory setting
Selection
Days (d), hours (h),
minutes (m) and
seconds (s)
Days (d), hours (h),
minutes (m) and
seconds (s)
• Cancel
Cancel
• Execute backup
• Restore
• Duplicate
• Compare
• Clear backup data
• Settings identical
Check not done
• Settings not
identical
• No backup
available
• Backup settings
corrupt
• Check not done
• Dataset
incompatible
91