Operating Tool - Endress+Hauser Proline Prosonic Flow B 200 Operating Instructions Manual

Ultrasonic transit time flowmeter
Hide thumbs Also See for Proline Prosonic Flow B 200:
Table of Contents

Advertisement

Proline Prosonic Flow B 200 HART
Endress+Hauser
The assignment of the measured variables to the dynamic variables can be modified and
assigned as desired via local operation and the operating tool using the following
parameters:
• Expert → Communication → HART output → Output → Assign PV
• Expert → Communication → HART output → Output → Assign SV
• Expert → Communication → HART output → Output → Assign TV
• Expert → Communication → HART output → Output → Assign QV
The following measured variables can be assigned to the dynamic variables:
Measured variables for PV (primary dynamic variable)
• Volume flow
• Corrected volume flow
• Corrected methane volume flow
• Energy flow
• Mass flow
• Methane fraction
• Gross calorific value
• Wobbe index
• Temperature
Measured variables for SV, TV, QV (secondary, tertiary and quaternary dynamic
variable)
• Volume flow
• Corrected volume flow
• Corrected methane volume flow
• Energy flow
• Mass flow
• Methane fraction
• Gross calorific value
• Wobbe index
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
9.3
Other settings
9.3.1
Burst mode functionality in accordance with HART 7
Specification
Navigation
"Expert" menu → Communication → HART output → Burst configuration → Burst
configuration 1 to 3
Burst configuration
Burst configuration 1 to 3
Burst mode 1 to 3
Burst command 1 to 3
System integration
53