Commissioning
Parameter
Current span
4 mA value
20 mA value
Failure mode
Failure value
64
Navigation
"Setup" menu → Current input
‣
Current input
Current span
4 mA value
20 mA value
Failure mode
Failure value
Parameter overview with brief description
Prerequsite
–
–
–
–
In the Failure mode
parameter, the Defined value
option is selected.
Description
Select current range for
process value output and
upper/lower level for alarm
signal.
Enter 4 mA value.
Enter 20 mA value.
Define input behavior in alarm
condition.
Enter value to be used by the
device if input value from
external device is missing.
Proline Prosonic Flow B 200 HART
Selection / User
Factory setting
entry
• 4...20 mA
Country-specific:
• 4...20 mA NAMUR
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA US
Positive floating-
700 mbar
point number
Positive floating-
Depends on country
point number
and nominal
diameter
• Alarm
Alarm
• Last valid value
• Defined value
Signed floating-point
0 mbar
number
Endress+Hauser
Need help?