Slow clock

If you have the analysis probe hooked up and running and observe a
slow clock or no activity from the interface board, the +5 V supply
coming from the analyzer may not be getting to the interface board.
❏ To check the +5 V supply coming from the analyzer, disconnect one of the
logic analyzer cables from the analysis probe and measure across pins 1
and 2 or pins 39 and 40.
• If +5 V is not present, check the internal analysis probe fuse or current
limiting circuit on the logic analyzer. For information on checking this fuse
or circuit, refer to the 1670G-Series Logic Analyzers Service Guide.
• If +5 V is present and the cable connection to the analysis probe appears
sound, contact your nearest Agilent Technologies Sales Office for
information on servicing the board.
Troubleshooting the Logic Analyzer
Analysis Probe Problems
427