To Set Up Time Correlation Between Two State Analyzers - Agilent Technologies 1670G Series User Manual

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

Using the Trigger Menu
Setting Up Time Correlation between Analyzers
To set up time correlation between two state
analyzers
To correlate the data between two state analyzers, both must have
Count Time turned on in their Trigger menus. Although both have
Count State available, it is not possible to correlate data based on
states even when they are identically defined.
1 In the Analyzer Trigger menu, select Count.
Count may be Count Off, Count Time, or Count States. Selecting the
field causes a pop-up to appear.
2 Select the field after Count: and select Time.
A warning may appear about reduced memory. It will not prevent you
from changing Count to Count Time.
3 Select Done.
4 Repeat steps 1 through 3 for the other state analyzer.
Now when you acquire data you will be able to interleave the listings.
82