The Scope Channel Menu; Offset Field - Agilent Technologies 1670G Series User Manual

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

The Scope Channel Menu

The Channel menu selects the channel input and the values that
control the vertical sensitivity, offset, probe attenuation factor, input
impedance, and coupling. The Channel menu also gives you preset
vertical sensitivity, offset, and trigger level values for ECL and TTL
logic levels. Each channel may be set independently of the other
channel.

Offset field

You use the Offset field to set the vertical position of the waveform on
the screen. Offset is the dc voltage that is added to or subtracted from
the input signal so that the waveform can be centered on the waveform
display. Offset range and resolution are dependent on vertical
sensitivity (V/Div) as shown in the table below. The table values are
based on a 1:1 probe setting.
Offset Range and Resolution
V/Div Setting
4 mV - 100 mV/Div
>100 mV - 500mV/Div
>500mV - 2.5 V/Div
>2.5 V - 10 V/Div
Offset changes are not reflected on the waveform until a Run is
initiated and the next acquisition is displayed. Changes to Offset during
a repetitive run will be seen on the next displayed acquisition because
the hardware is reprogrammed between acquisitions.
Using the Oscilloscope
The Scope Channel Menu
Offset Range
Offset Resolution
± 2V
1 mV
± 10V
1 mV
± 50V
1 mV
± 250V
2 mV
157

Advertisement

Table of Contents
loading

Table of Contents