Troubleshooting the LAN Connection
Troubleshooting the Initial Connection
Packets routinely lost
If packets are routinely lost, proceed to the troubleshooting section in
this chapter relating to your network.
Problems transferring or copying files
Copying files out of the logic analyzer
• If you have problems copying files out of the logic analyzer, you might be
experiencing timeout problems. See the timeout section on the previous
page.
• If you only receive 1 byte back when copying files, refer to "Dynamic Files"
later in this chapter.
Copying files into the logic analyzer
• If you have problems copying files into the logic analyzer, such as copying
setup or data to change a configuration, then check the File Timeout
setting in the LAN Settings menu. Refer to "Configure the network
addresses" in Chapter 1 for more information.
Communications not established
If you have just installed and configured the LAN interface and you
have never been able to read the logic analyzer remote file server
directory, go directly to the troubleshooting section relating to your
network in this chapter.
If you have been able to read the logic analyzer remote file server
directory and now cannot do so, check the following:
• Has any hardware been added or moved on your network? This includes
adding or removing any workstations or peripherals, or changing any
cabling.
• Have software applications been added to the network?