Agilent Technologies 1670G Series User Manual page 398

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

Logic Analyzer Concepts
The Trigger Sequence
For example, the following combinations are valid combinations for the
analyzer:
(a+b) • (In_Range2 + Timer2 > 400 ns)
(c • Out_Range1) + (f xor g)
The following combinations are not valid, because resources cross pair
boundaries:
a xor c
(d + Timer1 < 400 ns) • Edge1
398