1 Set up a timing analyzer. Go to the timing analyzer's Format
2 Go to the timing analyzer's Trigger menu.
3 Select an Edge term. Then assign glitch detection "*" to the
4 Go to the Arming Control menu. Set the Scope Arm In to
5 Select Group Run in the Analyzer Arm In menu.
6 Go to the Scope Trigger menu, and set Mode to Immediate.
NOTE:
To capture the waveform of a glitch using the
oscilloscope (oscilloscope option only)
The following setup uses the triggering capability of the timing
analyzer and the acquisition capability of the oscilloscope.
menu and set the Timing Acquisition Mode to Half Channel
500 MHz.
channels of interest represented by the Edge term.
Analyzer.
If you have trouble capturing the glitch waveform on the oscilloscope,
try adjusting the skew in the Arming Control menu, so the oscilloscope
triggers earlier.
A timing analyzer can trigger on a glitch and capture it, but a timing analyzer
doesn't have the voltage or timing resolution to display the glitch in detail. An
oscilloscope can display a glitch waveform with fine resolution, but cannot
trigger on glitches, combinations of glitches, or sophisticated patterns
involving many channels.
Triggering Examples
Cross-Arming Trigger Examples
123