Pattern/Edge
Combinations
Time Violations
Delay
1. Find edge and pattern
This function becomes true when a selected edge type is seen at the
same time as a designated pattern. It uses one internal sequence level.
2. Find pattern occurring too soon after edge
This function becomes true when a designated pattern is seen
occurring within a set duration after a selected edge type is seen. It
uses three or four internal sequence levels.
3. Find pattern occurring too late after edge
This function becomes true when one selected edge type occurs, and
for a designated period of time after that first edge is seen, a pattern is
not seen. It uses two internal sequence levels.
1. Find 2 edges too close together
This function becomes true when a second selected edge is seen
occurring within a designated period of time after the occurrence of a
first selected edge. It uses three or four internal sequence levels.
2. Find 2 edges too far apart
This function becomes true when a second selected edge occurs
beyond a designated period of time after the first selected edge. It uses
two internal sequence levels.
3. Find width violation on a pattern/pulse
This function becomes true when the width of a pattern violates
designated minimum and maximum width settings. It uses four or five
internal sequence levels.
• Wait "t" seconds
This function becomes true after a designated time period has expired.
It uses one internal sequence level.
Logic Analyzer Reference