Fidas24 - Substitute Gas Calibration - ABB EasyLine EL3000 Operating Instruction

Continuous gas analyzers
Hide thumbs Also See for EasyLine EL3000:
Table of Contents

Advertisement

EL3000 CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D
Test gas connection at the sample gas inlet during manual
calibration
If the test gas is connected directly at the sample gas inlet
during manual calibration, when the pressure switch is activated
the following message appears in the LCD display:
'Error. Pressure switch has not detected any calibration gas!'
Check test gas flow and repeat calibration with .
This message can be bypassed by pressing  ESC.
Wait until the warm-up phase has ended
Analyzer module may not be calibrated until the warm-up phase
has ended.
Fidas24 – Substitute gas calibration
Example
Substitute gas calibration in the Fidas24 is described using the
'Acetone measurement in room air' example.
Acetone measurement in room air
Acetone (C
H
O) in higher concentrations cannot be filled in test
3
6
gas containers.
For this reason, the analyzer module is factory-set for calibration
with the substitute gas propane (C
Measuring ranges
Component 1
Measuring range 1
C
H
O in room air
3
6
Component 2
Measuring range 1
C
H
in N
(substitute gas)
3
8
2
Calibration data
Calibration method
Substitute gas calibration
Zero component
Component 1
C
H
3
6
Component 2
C
H
3
8
Span component
Component 1
C
H
3
6
Component 2
C
H
3
8
Calibration
Calibrate zero point with air (Component 1 or Component 2).
Calibrate end point with test gas propane in N
Other measurement tasks
For other measurement tasks select the test gases and
measurement ranges in a similar manner according to the
sample gas composition.
H
) in N
.
3
8
2
0 to 10,000 ppm
0 to 10,000 ppm
Measuring range 1 or
O
Measuring range 1
Measuring range 1 or
O
Measuring range 1
.
2
137