External Signal Input (Optional, Only Ext And Add); General Function - NF DP015S Instruction Manual

Dp series programmable ac power source
Hide thumbs Also See for DP015S:
Table of Contents

Advertisement

10.22.3 External Signal Input (Optional, only EXT and ADD)

This multiplies the input signal by the specified gain and outputs it. For ADD, the internal signal
source is added.
EXT: Output voltage (V) = External signal input (V) x Gain (V/V)
ADD: Output voltage (V) = External signal input (V) x Gain (V/V) + Internal signal source setting
(V)
Setting Range for gain
Setting resolution
Gain accuracy
Input-output phase
Input terminal
Input impedance
Input voltage range
Nondestructive
maximum input
voltage
Input frequency range
Note: There is no optional setting for the polyphase model. It cannot be used for the polyphase
system or polyphase output of the Multi-phase.

10.23 General Function

LCD
Contrast
display setting Color
Beep
(key operation,
erroneous operation)
Key lock
Output relay control
Output setting
at power-on
Trigger output setting
Time unit setting for
Sequence and Simulation
Reset function
DP Series
Specification
100 V range: 0.0 to 220.0 times
200 V range: 0.0 to 440.0 times
(DC, 45 Hz to 65 Hz, the gain is the factory default, the rated output
BNC connector (rear panel, unbalanced)
Also used as the external synchronization signal input
±2.2 V (A/D resolution 10 bit)
DC to 550 Hz (sine wave)
DC to 100 Hz (other than sine wave)
Description
Blue tone or white tone
Alarms on abnormal situation regardless of the setting
On: Only key lock-off and output-off are available
On: The output relay is used to turn the output on/off
Off: The output relay is not used. High impedance to
turn the output off
On: Output on after power-on
Polarity: positive
Pulse width: 0.1 ms to 10 ms (resolution 0.1 ms)
Resets the items stored in the System Setting Memory (excluding the
external interface setting) and the items that are to be reset at
power-on, to the factory default settings.
0.1
±5 %
voltage, no load)
In-phase
1 MΩ
±10 V
0 to 99
On or Off
On or Off
On or Off
or negative
ms or s
10.23 General Function
Factory
default
100
200
Factory
default
Blue tone
On
Off
On
Off
Negative
10 ms
s
263

Hide quick links:

Advertisement

Table of Contents
loading