Agc (Optional); Autocal (Output Voltage Compensation) - NF DP015S Instruction Manual

Dp series programmable ac power source
Hide thumbs Also See for DP015S:
Table of Contents

Advertisement

10. Specifications

10.16 AGC (Optional)

When the AGC (Automatic Gain Control) is on, the detection point voltage is always measured, and
the output voltage is continuously corrected so that its effective value is equal to the output voltage
setting value. The fluctuation of the detection point voltage can be suppressed even when the load is
fluctuated. The detection point can be switched between the sensing input terminal (remote sensing
on) and the output terminal (remote sensing off).
Effective only for AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA and when the waveform
is sine wave or DC. It is turned off when the Sequence or Simulation is selected. It cannot be
selected when the Autocal is set to on.
Response
time
Operation
range
Calibration
range
The output voltage should be within the allowed voltage setting range of the product.
Accuracy

10.17 Autocal (Output Voltage Compensation)

When the Autocal (Automatic Calibration) is on, the detection point voltage is always measured,
and the output voltage is continuously corrected so that its effective value is equal to the output
voltage setting value. The ratio (correction factor) of the detection point voltage to the output
voltage setting value is used until the Autocal or the power is turned off. Therefore, the detection
point voltage is not necessarily maintained if the load changes while the Autocal is on. The detection
point can be switched between the sensing input terminal (remote sensing on) and the output
terminal (remote sensing off).
Unlike the AGC, it cannot follow a load fluctuation because it does not keep track of the voltage. On
the other hand, when the load is stable, it has a merit of short response time on changing the output
voltage setting.
Effective only for AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA and when the waveform
is sine wave or DC. It is turned off when the Sequence or Simulation is selected. It cannot be
selected when the AGC is set to on.
Restriction
when on
Calibration
range
The output voltage should be within the allowed voltage setting range of the product.
Accuracy
Note: The values of the calibration range and accuracy are the ones at the time when the Autocal
is turned on.
258
(DC/50 Hz/60 Hz, at the rated output voltage)
The output voltage setting is 50 V or higher
Within ±10% (difference between the output voltage and measured value)
(in the case of DC, 40 Hz to 550 Hz, 50 V or higher output voltage,
resistance load, the output current is the maximum current or less)
The output voltage setting is 50 V or higher
Within ±10% (difference between the output voltage and measured value)
(in the case of DC, 40 Hz to 550 Hz, 50 V or higher output voltage,
resistance load, the output current is the maximum current or less)
All models
Within 100 ms (typ.)
Within ±0.5 V / ±1.0 V
All models
Within ±0.5 V / ±1.0 V
PROGRAMMABLE AC POWER SOURCE

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents