10.22.3 External Signal Input (only EXT and ADD)

This multiplies the input signal by the specified gain and outputs it. For ADD, the internal signal
source is added.
EXT: Output voltage (V) = External signal input (V)  Gain (V/V)
ADD: Output voltage (V) = External signal input (V)  Gain (V/V)
Setting Range for gain
Setting resolution
Gain accuracy
Input-output phase
Input terminal
Input impedance
Input voltage range
Nondestructive
maximum input voltage
Input frequency range
*71:
Not available for the polyphase model. It cannot be used for the polyphase system or polyphase
output of the Multi-phase.
*72:
DC or 45 Hz to 65 Hz, the gain is the factory default, the rated output voltage, no load.

10.23 General Function

LCD display
Contrast
setting
Color
Beep (key operation,
erroneous operation)
Key lock
Output relay control
Output setting
at power-on
Trigger output setting
Time unit setting for
Sequence and Simulation
Reset function
DP Series/DP Series Type R
100 V range: 0.0 to 227.0 times
200 V range: 0.0 to 454.0 times
0.1
*72 5 %
In-phase
BNC connector (rear panel, unbalanced)
Also used as the external synchronization signal input
1 M
2.2 V (A/D resolution 10 bit)
10 V
DC to 550 Hz (sine wave)
DC to 100 Hz (other than sine wave)
0 to 99
Blue tone or white tone
On or Off
Alarms on abnormal situation regardless of the setting
On or Off
On: Only key lock-off and output-off are available
Enable: The output relay is used to turn the output
on/off
Disable: The output relay is not used. High impedance
to turn the output off
On or Off
On: Output on after power-on
Polarity: positive
Pulse width: 0.1 ms to 10 ms (resolution 0.1 ms)
ms or s
Resets the items stored in the System Setting Memory (excluding the
external interface setting and external control setting) and the items that
are to be reset at power-on, to the factory default settings.
+ Internal signal source setting (V)
Specification *71
Description
or negative
10.23 General Function
Factory default
100
200
Factory default
Blue tone
On
Off
Enable
Off
Negative
10 ms
s
293