Using Frequency Setting Range Limit - NF DP015S Instruction Manual

Dp series programmable ac power source
Hide thumbs Also See for DP015S:
Table of Contents

Advertisement

2. The screen for the effective value setting or the peak value setting opens depending on the
signal source setting/output waveform settings made here. Set the voltage setting range limiting
value in the [V] item (effective value setting) and [+Vop]/[ − Vop] item (peak value setting).
Effective Value Setting
3. Close the window.

4.1.4 Using Frequency Setting Range Limit

You can limit the range of the frequency setting value for the internal signal source by using the
frequency setting range limit. Even if trying to set a frequency that exceeds the limit, the setting
value is restricted to the limiting value. This can prevent mis-setting when you do not want to set
values outside the range.
The frequency setting range limiting value is common to both 100 V and 200 V ranges.
For the default setting range, see 10.14.
-------- Notes----------------------------------------------------------------------------------------------------------------
 The frequency setting range limit is useful for the prevention of an unintended frequency
setting when a setting value is rapidly increased or decreased by using the shuttle.
---------------------------------------------------------------------------------------------------------------------------------
DP Series
4.1 Using Limiter, Setting Limiter
Peak value setting
73

Hide quick links:

Advertisement

Table of Contents
loading